Operation
Eppendorf ThermoMixer
®
C
English (EN)
23
SmartBlock plates, SmartBlock PCR 96 or SmartBlock PCR 384:
5.5 Installing the ThermoTop
The ThermoTop is compatible with thermoblocks which feature the condens.protect
symbol:
The condens.protect technology available with ThermoTop prevents the formation of
condensation on the inner wall or the lid of the tube.
Prerequisites
• A compatible thermoblock has been attached.
• Tubes or plates have been inserted.
1 Height sensor
To ensure uniform temperature control,
place the lid on the thermoblock.
Place the ThermoTop on the device
vertically from above. The centering pins
behind the heating/cooling plate fit into
the recesses of the ThermoTop.
• The ThermoTop is correctly positioned if
the seal is fully flush with the upper part
of the device.
• The blue LED of the ThermoTop lights.
• The symbol appears in the display.