Instructions for use
Rotor A-4-81, Rotor A-4-81-MTP/Flex
English (EN)
Your local distributor: www.eppendorf.com/contact
Eppendorf AG · 22331 Hamburg · Germany
eppendorf@eppendorf.com · www.eppendorf.com
Eppendorf
®
and the Eppendorf logo are registered trademarks of Eppendorf AG.
5810 718.961-14/042013
3.2.1 Loading the rectangular bucket
Insert the adapters into the rectangular bucket with the black clamp facing the
"Eppendorf" logo on the bucket.
For tubes with a length of more than 100 mm perform a manual swing-out test with
empty tubes.
The tubes must not come into contact with the rotor cross.
3.2.2 Loading the plate buckets
Observe the maximum load and the maximum height when using stacked plates.
4 Maintenance
4.1 Cleaning
• All buckets, buckets, caps and adapters can be autoclaved (121°C, 20 min).
• The rotor cross is not autoclavable.
Clean the rotor and accessories with a mild cleaning agent.
5 Technical data
Rectangular bucket
500 mL
Bucket for concical
tubes 7 × 50 mL
MTP/Flex bucket
Max. g-force (rcf) 3 220 × g 3 220 × g 2 900 × g
Max. speed (rpm) 4 000 rpm 4 000 rpm 4 000 rpm
Max. load per bucket
(adapter, vessel and
content)
780 g 7 × 75 g 380 g
Max. centrifugation
radius
18 cm 18 cm 16.3 cm
Max. loading height/
vessel length
Please note the data in the operating
manual of the centrifuge.
60 mm