Eppendorf ThermoMixer C Bedienungsanleitung Seite 36

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken
  • Seite
    / 54
  • Inhaltsverzeichnis
  • LESEZEICHEN
  • Bewertet. / 5. Basierend auf Kundenbewertungen
Seitenansicht 35
36
Troubleshooting
Eppendorf ThermoMixer
®
C
English (EN)
6 Troubleshooting
If you cannot remedy an error with the recommended measures, please contact your local
Eppendorf partner. The contact address can be found online at: www.eppendorf.com
.
6.1 General errors
Problem Cause Solution
Display remains
dark.
No mains connection.
Check the mains connection
and the power supply.
Switch the device on.
Set temperature
is not reached.
Set temperature is more than
15 °C below ambient temperature.
Set up the device in a cooler
environment.
ThermoTop LED
does not light.
No thermoblock has been
attached
The thermoblock is not
compatible with ThermoTop.
Use a compatible thermoblock
with a condens.protect symbol:
The interface between the
device and the ThermoTop is
dirty.
Remove any dirt from the front
of the ThermoTop.
Remove any dirt from the top of
the device, especially from the
viewing window beside the
heating/cooling plate.
ThermoTop does
not fit on the
device.
The thermoblock is not
compatible with ThermoTop.
The lid is attached to the
thermoblock.
Use a compatible thermoblock
with a condens.protect symbol:
If using the ThermoTop, do not
use the lid.
The device does
not mix nor
control the
temperature.
Various causes are possible.
Contact your local Eppendorf
partner.
Seitenansicht 35