Operation
Eppendorf BioSpectrometer
®
basic
English (EN)
26
5.3.2.2 Checking parameters
5.3.2.3 Measuring the blank and standards
Check the parameter setting. The [Page dn] and
[Page up] softkeys allow you to call up the
parameter list pages. You can modify and save
parameters using [Edit].
For evaluations without standards (e.g. DNA measurements), this method step is omitted.
1. Start by measuring a blank (blank key).
2. Then measure all standards one by one
(standard key).
The display always marks the standard that is to be
measured next. Use the [Graph] resp. [Table]
softkey to change the result view.
Press [Next] to accept the evaluation calculated
from the standard results.