Error Cause Solution
Droplets on
the inner wall
of the pipette
tip.
Uneven wetting of the
plastic wall.
A pipette tip with poor
wetting properties has
been used.
Attach a new pipette tip.
Use an original eppendorf
tip.
Pipette is
dripping
and/or the
volume
pipetted is
incorrect.
The tip is loose.
A poorly fitting pipette tip
has been used.
Liquid with a high vapor
pressure has been pipetted.
Tip has been taken out of
the liquid too quickly.
The pipette is dripping
because:
Piston is contaminated.
Piston is damaged.
Seals are damaged.
Tip cone loose.
Press the tip firmly in place.
Use an original eppendorf
tip.
In this case, we recommend
pipetting using a positive-
displacement pipette.
Remove the tip slowly from
the liquid.
Clean and lightly lubricate
the piston.
Replace the piston and seal
(see Part B, Maintenance).
Replace all seals
(see Part B, Maintenance).
Lightly tighten tip cone with
wrench (see Part B,
Maintenance).
Exchange, where
necessary.
10 Troubleshooting
Eppendorf Reference® – Part A – Chapter 10