Operation
InjectMan
®
4
English (EN)
60
6.8.2 Deleting the lower limit
1. Press Z-axis Limit.
The limit is deleted.
6.8.3 Defining the upper limit
1. In the Installation menu, select the Upper limit submenu.
The Upper Limit window appears.
2. Move the capillary to the top position.
3. Save the position with Set.
The Z-coordinate is marked with U.
The value for the limit (Lim) is displayed.
The capillary cannot be moved any higher.
6.8.4 Deleting the upper limit
1. In the Installation menu, select the Upper limit submenu.
The Upper Limit window appears.
2. Save the position with Clear.
The limit is deleted.