6.12 Step injection function
With this function you can execute a straight injection over a defined distance. Step
injection can only be executed when a FemtoJet is connected.
6.12.1 Step injection menu and parameters
Abb. 6-3:Menu – Step injection
Fig. 6-3: Menu – Step injection
Parameter Value Range of values Increment Standard
Step injection Switch the function on or off OFF/ON OFF
Step inj. speed Set the injection speed in μm
per second
5 – 10000 5 300
Step inj. dist. Set the injection path in μm 1 – 2000 1 20
Step inj. axial Switch the axial movement of
the capillary on or off
OFF/ON On
Step inj. back Switch the backward
movement of the capillary on
or off
OFF/ON On
Synchr. inject. Set the time of injection MOVE
IMMEDIATE
LIMIT
PRESSURE
LIMIT
Parameter Range of values Function
Synchr. inject. MOVE Injection movement without injection pressure
IMMEDIATE Injection pressure starts with the injection movement
LIMIT Injection pressure at the end of the injection distance
(Step inj. dist.)
PRESSURE Injection pressure without injection movement