7 Troubleshooting
7.1 General errors
7.1.1 Motor module
7.1.2 Capillary
7.1.3 Control board and display
Symptom/message Cause Remedy
Motor axes move in the
wrong direction or do
not match joystick
movement.
Motor module mounted
incorrectly.
Incorrect installation
parameters entered.
Direction of movement of
axis inverted.
Compare the installation
parameters with the
structure of the motor
module.
Check the orientation and
structure of the modules.
Check the connections of
the modules at the control
board.
Cancel the inversion of the
axis.
Symptom/message Cause Remedy
Capillary moves too
fast or too slow.
Radius of working range
incorrectly set.
Set the radius using the
selection dial or in the Speed
menu.
Incorrect acceleration
factor.
In the Installation menu, set
the value for the Dyn-factor
parameter.
Capillary only moves to
the side or vertically.
Y-axis is disabled.
The Y off function disabled.
Capillary does not
move down far
enough.
•The Z-axis Limit function is
active.
Capillary is incorrectly
adjusted.
Deactivate the Z-axis Limit
function.
Readjust the capillary.
Symptom/message Cause Remedy
Device does not
respond to keystroke
when Home function is
active.
Function is active.
Press the home key again.
The capillary moves down.
Press the Back manual
softkey.
Move the joystick.